CWA- Alk-T01 Total Alkalinity Analyzer

Category

CWA-Alk-T01 is an online total alkalinity analyzer that measures the total alkalinity of water using titration and colorimetric methods. It automatically pumps reagents into the water sample for reaction, determines the reaction endpoint via colorimetry, and then calculates the total alkalinity based on the consumed reagent volume. By selecting different reagents, the analyzer can measure various types of total alkalinity.

  • Accurate and reliable fully automatic alkalinity analyzer.
  • With self-diagnosis, abnormal alarm, and data exceeding the standard alarm function
  • Interval measurement optional, minimum 5min
  • Supports Modbus RS485 and 4-20mA signal outputs
  • Support 4G, WiFi wireless data transmission

★  Key Performance Parameters:

 

Monitoring parameter Range Accuracy Repeatability
Total Alkalinity 2.5-2050 mg/L(ppm) CaCO₃ ±5 % F.S. ±5 % F.S.

 

Reagent range

No. Reagent type mg/L(ppm) CaCO3 mmol/L
6400 WAR-Alk-T00 2.50-45 0.05-0.90
6401 WAR-Alk-T01 5.34-134 0.107-2.68
6402 WAR-Alk-T02 8.01-205 0.160-4.10
6403 WAR-Alk-T03 10.7-267 0.214-5.34
6404 WAR-Alk-T04 16.0-401 0.32-8.02
6405 WAR-Alk-T05 80.1-2050 0.160-41

 

 

Specification:

Additional information

Measurement Method

Titrimetric colorimetric method

Dimension(H x W x D)

300×300×180 mm

Weight

Approx. 4 kg

Power Requiremens

100-240 VAC, 50/60 Hz

Power Consumption

25W

Operating Temperature

5 – 45 °C (41 – 113 °F)

Storage Temperature

5 – 45 °C (41 – 113 °F)

Protection Class

IP54

Humidity

20 % – 95 %, indoor installation

Pressure

0.2 – 2 bar

Tubing Requirements

Inlet/Outlet: 6 mm outer diameter

Communication

RS 485 (Modbus RTU), WiFi, 4G (optional)

Analog Output

0/4 – 20 mA, Resolution: < 100 μA, Max. load: 750 Ω

Digital Input

1~way I/O input (for start analysis/ flow switch)

Digital Output

2 relays (max. 250 VAC / 30 VDC 3A)

Data Storage

32G

Cable Length

6 m (6.56 ft), Customizable length

Warranty

One year